Menu
Back

HC2-IExxx-M

Stainless Steel Probe for Pressurized Conduits and Vessels

The metal industrial probe is especially suitable for high temperatures, demanding industrial environments and areas where hygiene plays an important role. The probe measures humidity and temperature and calculates the dew/frost point.

Applications
Food and pharmaceutical production, drying processes, industrial manufacturing

Features
• Range of application: -100…200 °C, (screw-in probe; -50…200 °C) / 0…100 %rh
• Digital interface (UART) and scalable analog outputs, 0…1 V
• Standard output scaling: 0…1 V = -40…60 °C / 0…100 %rh
• Adjusted at 23 °C and 10, 35, 80 %rh

Probe type Screw-in probe
Weight 290 g
Operating range -100…200 °C / 0…100 %rh
Accuracy at 23 ±5 °C ±0.8 %rh / ±0.1 K
Accuracy w. adjustment profile "standard" at 23 °C / 10, 35, 80 %rh ± 0.8 %rh / ± 0.1 K
Power supply / Power consumption 3.3 V ± 0.1V / ~4.5 mA
Long-term stability <1 %rh / year
Humidity sensor Hygromer IN-1
Temperature sensor PT100 1/3 DIN Class B
Probe protection Filter carrier, brass nickel plated
Filter cartridge depending on the application, order separately
Pressure Pressure-resistant to 100 bar / 1450 PSI
Response time t 63 <15 sec.
Allowed wind speed 3 m/s
Psychrometric calculations Dew or frost point
Connection thread of probe 1/2" G / 1/2" NPT
Reproducibility Airchip <0.02 %rh / 0.01 K
Exploitation software ROTRONIC HW4
Probe adjustment by software 1 Point & multipoint %rh & °C
Firmware update With internet and HW4 software
User information from Software HW4 (devicemanager)
Data processing via HW4 with interface cables
Device Protection by password Yes
Type of output signals 2 x 0…1 VDC
Analogue outputs scaleable by the user Yes
Analog output signal (standard) 0…1 V = 0…100 %rh; 0…1 V = -40…60 °C
Digital output signal (standard) UART
Service Interface UART
Reverse voltage protection Mechanically in the connector
Max. cable length Up to 5 m with passive cable. Up to 100 m with active cable.
Probe material Stainless steel 1.4305 (V2A)
Probe cable material FEP, black
Enclosure material Stainless steel 1.4305 (V2A)
FDA/GAMP compatibility FDA 21 CFR Part 11 and GAMP compatible