Probe Calibration
The digital HygroClip probes can be calibrated and adjusted at one or more points with the HygroLab indicator:

• For a quick check of the probe and eventual 1-point adjustment of humidity and temperature, you can plug a
  freshly calibrated probe into any free probe input and adjust against this probe.

• For scheduled maintenance, you can adjust temperature at up to 2 points and humidity at up to 4 point against
  physical references.

You may also make use of the full interchangeability of the HygroClip digital probes and rotate the probes with a calibration laboratory. The calibration laboratory can do any of the following:

  1. Use the HygroLab 2 indicator to directly calibrate and adjust the probe
  2. Use any other device such as the HygroClip DI network interface and adjust the probe with a PC on which the HW4 software has been installed
  HW4 Software: software based probe calibration is both simple and precise: there are no hard-to-reach, hard-to-adjust potentiometers. HW4 offers a high level of flexibility in selecting the calibration reference, giving the user complete control over the calibration process. HW4 features the following calibration modes:

• Humidity: 1-point (offset) or multi-point (up to 4 points)
• Temperature: 1-point (offset) or 2-point (offset and slope)

When humidity calibration is done using the ROTRONIC humidity standards, HW4 automatically corrects the value of the standard to reflect the temperature of calibration (see note). HW4 also offers the possibility of printing a certificate of calibration.

Note: the value of the standards does not depend on barometric pressure (altitude).

Calibration standards 

Certified Humidity Standards:
if you do not have access to a humidity generator, the ROTRONIC humidity standards offer a reliable and convenient way to generate a reference humidity condition during probe calibration. The ROTRONIC standards are shipped with an internationally accepted certificate, meeting in full the requirements of quality systems such as ISO9000. The standards are available in boxes of 5 glass ampoules of the same value, and can be stored indefinitely. Standards in the range of 5 to 95 %RH are non-saturated aqueous salt solutions. The 0 %RH humidity standard is made of small granules of a highly porous ceramic. A Material Safety Data Sheet is available for each standard.

Calibration devices 

Calibration Devices:
some of the probes used with the HygroLab 2 may require a calibration device to hold the humidity standard. This device is a small airtight container that fits on the probe and seals around the humidity sensor. During calibration, a known reference humidity is produced inside the calibration device by means of a humidity standard.

Four calibration devices are available for the HygroClip probes:
ER-15 for 15mm diameter probes
ER-05 for 5mm diameter probes

ORDERING INFORMATION
Order code: Description: Order code: Description:
HW4-P-QUICK HW4 software package EA00-SCS 0.5%RH cal. std, SCS Certificate, pack 5
    EA05-SCS 5%RH cal. std, SCS Certificate, pack 5
    EA10-SCS 10%RH cal. std, SCS Certificate, pack 5
    EA11-SCS 11.3%RH cal. std, SCS Certificate, pack 5
    EA20-SCS 20%RH cal. std, SCS Certificate, pack 5
    EA35-SCS 35%RH cal. std, SCS Certificate, pack 5
    EA50-SCS 50%RH cal. std, SCS Certificate, pack 5
ER-15 Calibration device for 15mm diameter probes EA65-SCS 65%RH cal. std, SCS Certificate, pack 5
ER-05 Calibration device for 5mm diameter probes EA75-SCS 75.3%RH cal. std, SCS Certificate, pack 5
    EA80-SCS 80%RH cal. std, SCS Certificate, pack 5
    EA95-SCS 95%RH cal. std, SCS Certificate, pack 5