| Probe Calibration |
|
| The
digital HygroClip probes can be calibrated and
adjusted at one or more points with the HygroLab indicator:
For a quick check of the probe and eventual 1-point adjustment of
humidity and temperature, you
can plug a |
|
|
|
|
HW4 Software:
software based probe calibration is
both simple and precise: there are no hard-to-reach,
hard-to-adjust potentiometers. HW4 offers a high level of flexibility in
selecting the calibration reference, giving the user complete control over
the calibration process. HW4 features the
following calibration modes: Humidity: 1-point (offset) or multi-point (up to 4 points) Temperature: 1-point (offset) or 2-point (offset and slope) When humidity calibration is done using the ROTRONIC humidity standards, HW4 automatically corrects the value of the standard to reflect the temperature of calibration (see note). HW4 also offers the possibility of printing a certificate of calibration. Note: the value of the standards does not depend on barometric pressure (altitude). |
|
Certified Humidity Standards: if you do not have access to a humidity generator, the ROTRONIC humidity standards offer a reliable and convenient way to generate a reference humidity condition during probe calibration. The ROTRONIC standards are shipped with an internationally accepted certificate, meeting in full the requirements of quality systems such as ISO9000. The standards are available in boxes of 5 glass ampoules of the same value, and can be stored indefinitely. Standards in the range of 5 to 95 %RH are non-saturated aqueous salt solutions. The 0 %RH humidity standard is made of small granules of a highly porous ceramic. A Material Safety Data Sheet is available for each standard. |
|
Calibration Devices: some of the probes used with the HygroLab 2 may require a calibration device to hold the humidity standard. This device is a small airtight container that fits on the probe and seals around the humidity sensor. During calibration, a known reference humidity is produced inside the calibration device by means of a humidity standard. Four calibration devices are available for the HygroClip probes: ER-15 for 15mm diameter probes ER-05 for 5mm diameter probes |
|
||||||||||||||||||||||||||||||||||||||||||||||||||||