Home Page | Back to HygroGen Main Page
| HygroGen Overview |
|
![]() |
The HygroGen humidity
temperature generator uses a mixed flow
method for generating the humidity (%RH) required by the user. A desiccant
cartridge provides low humidity, a saturator high humidity, and a control
system mixes these together to create a constant environment within a well
insulated chamber. Temperature is controlled with a Peltier thermoelectric
heat pump. Set points are easily adjusted using the front panel controller
keypad, or via the RS-232 interface using a PC and optional controller
software. The
HygroGen uses a high performance controller to achieve both exceptional
stability and rapid stabilization when moving from one condition to
another. Smart chamber design, combined with a high rate of air flow, contributes
to minimizing humidity and temperature gradients inside the chamber,
and to shielding the chamber from the outside environment. |
|
|
For the
purpose of controlling conditions inside the chamber, humidity is measured
with a HygroClip S digital probe and temperature with a separate Pt100
RTD. Depending on the model of HygroGen, the reference humidity and
temperature used to calibrate humidity instrumentation is usually provided by
another set of probes and instruments |
||
| Key Features | ||
![]() |
1 | Dryer cell - removable cartridge filled with a molecular sieve desiccant (can be regenerated) |
| 2 | Distilled water tank. Fill or empty by inserting the dosing syringe in the access port located on the front of the unit | |
| 3 | Multi-loop controller especially configured and tuned for the HygroGen | |
| 4 | Internal chamber designed to provide a stable calibration environment with minimal temperature gradients. | |
| 5 | Chamber door - different models available | |
| 6 | Stainless steel enclosure - facilitates cleaning | |