HygroGen
Specifications
Control range maximum 5...95%RH and 5...50°C
Control range minimum 10...90%RH over the range of 10...50°C
Humidity control stability
(peak to peak)
0.02%RH at 5%RH to 0.35%RH at 95%RH and 20°C
0.11 %RH at 15 %RH and 45°C
0.50 %RH at 10%RH to 0.35%RH at 90 %RH and 5°C
Temperature control stability
(peak to peak)
0.03°C or better from 5 to 45°C (at any humidity value)
Temperature gradient limit 1 0.10°C or better from 20°C to 45°C and 15 %RH to 80 %RH
Humidity gradient limit 1, 2 0.2 %RH or better from 5 to 80%RH
1.7 %RH or better at 95 %RH
Time to set point 35 to 80 %RH change at 20°C: 5 minutes or better
20 to 45°C change: 20 minutes or better
Temperature control probe Pt100 RTD 1/3 Class B, directly wired to controller
Humidity control probe ROTRONIC HygroClip S

Chamber volume

2 liters / 122 cu ft

Overall dimensions

455 x 420 x 212 mm / 17.92 x 16.54 x 8.35" 

Weight

17 kg / 37.5 lb

Power

110-230 VAC 3A  50-60Hz  Class I


1 at a probe immersion depth of 120 mm / 4.7", measured from the back of the door
2 at
20°C
 
HygroGen Humidity Temperature Map

1:   Range of operation

2:   Hatched regions indicate conditions when condensation can be expected in the
chamber