| HygroGen Specifications |
|
| Control range maximum | 5...95%RH and 5...50°C |
| Control range minimum | 10...90%RH over the range of 10...50°C |
|
Humidity control stability (peak to peak) |
0.02%RH at 5%RH to 0.35%RH
at 95%RH and 20°C 0.11 %RH at 15 %RH and 45°C 0.50 %RH at 10%RH to 0.35%RH at 90 %RH and 5°C |
|
Temperature control stability (peak to peak) |
0.03°C or better from 5 to 45°C (at any humidity value) |
| Temperature gradient limit 1 | 0.10°C or better from 20°C to 45°C and 15 %RH to 80 %RH |
| Humidity gradient limit 1, 2 |
0.2 %RH or better from 5 to 80%RH 1.7 %RH or better at 95 %RH |
| Time to set point |
35 to 80 %RH change at 20°C: 5 minutes or
better 20 to 45°C change: 20 minutes or better |
| Temperature control probe | Pt100 RTD 1/3 Class B, directly wired to controller |
| Humidity control probe | ROTRONIC HygroClip S |
|
Chamber volume |
2 liters / 122 cu ft |
|
Overall dimensions |
455 x 420 x 212 mm / 17.92 x 16.54 x 8.35" |
|
Weight |
17 kg / 37.5 lb |
|
Power |
110-230 VAC 3A 50-60Hz Class I |
|
1 at a probe immersion depth of 120 mm / 4.7", measured from the back of the door 2 at 20°C HygroGen Humidity Temperature Map |
|||
|
|||