Back to HygroClip S3
HygroClip S3 Humidity Temperature Probe
Technical data Humidity sensor ROTRONIC Hygromer® C94 (capacitive) Temperature sensor RTD Pt100 Ohm, 1/3 DIN Sensor protection Wire mesh filter (standard) / foam filter (optional) Measuring range 0...100%RH and -40...85°C ( -40...185°F) Operating limits same as measuring range Accuracy at 23°C / 73°F ±1.5%RH and ±0.2°C / 0.4°F (when calibrated for HygroFlex M) Repeatability better than 0.5%RH and 0.1°C/°F Analog output signals
(minimum load > 10 k Ohm)0...100%RH = 0...1V
- 40...85°C = 0...1.25V (-40...60°C = 0...1V)Digital output/input signal 1-wire, HygroClip format, 2-way communication Analog signal resolution 0.02%RH and 0.06°C Digital signal resolution 0.004%RH and 0.004°C Sampling time < 0.7 sec Maximum cable length 5 m /15 ft for direct output
up to 100m / 325 ft with signal amplifierConnection type bayonet cap on mounting connector (see accessories) Protection grade IP65 (probe to connector) Supply Voltage 3.5...50VDC Current Consumption < 4mA Minimum excitation time < 4 sec EMC compatibility (CE) EN-50081-2, EN50082-2 Material Polycarbonate, white Dimensions length: 100mm (3.94"), diameter: 15mm (0.59") Weight 14 g / 0.5 oz