Home Page
| HygroClip
S3 Main Page
HygroClip S3 Probe Calibration
| See also: HW3 System Requirements | |
| The HygroClip S probe can be calibrated and adjusted at one or more points with the HW3 software after connecting the probe to a PC with the MOK-03-WIN cable (see accessories): | |
|
HW3 Software:
software based probe calibration is
both simple and precise: there are no hard-to-reach, hard-to-adjust
potentiometers. HW3 offers a high level of flexibility in selecting the
calibration reference, giving the user complete control over the
calibration process. HW3 features the following calibration modes: • Humidity: 2, 3 or 4 points • Temperature: 2 points • 1-point adjustment function (offset) for both humidity and temperature When humidity calibration is done using the ROTRONIC humidity standards, HW3 automatically corrects the value of the standard to reflect the temperature of calibration (see note). HW3 also offers the possibility of printing a certificate of calibration. Note: the value of the standards does not depend on barometric pressure (altitude). |
|
Certified Humidity Standards: if you do not have access to a humidity generator, the ROTRONIC humidity standards offer a reliable and convenient way to generate a reference humidity condition during probe calibration. The ROTRONIC standards are shipped with an internationally accepted certificate, meeting in full the requirements of quality systems such as ISO 9000. The standards are available in boxes of 5 glass ampoules of the same value, and can be stored indefinitely. Standards in the range of 5 to 95 %RH are non-saturated aqueous salt solutions. The 0 %RH humidity standard is made of small granules of a highly porous ceramic. A Material Safety Data Sheet is available for each standard. |
|
Calibration Devices: the ER-15 calibration device is a small airtight container that fits on the HygroClip S probe and seals around the humidity sensor. During calibration, a known reference humidity is produced inside the calibration device by means of a humidity standard. |
|
||||||||||||||||||||||||||||||||||||||||||||||||||||