Each M100 series
transmitter features a service connector that can be accessed without
disturbing the output signals. Cable ACRLXB5 is used to connect the service connector to the HygroPalm 3 portable indicator. |
|
Field Calibration b) single- or multi-point
adjustment against a reference environment (requires a humidity
generator). When no humidity generator is available, ROTRONIC offers
easy-to-use certified humidity standards to generate known humidity values.
See below for details.
|
![]() |
Certified Humidity Standards: if you do not have access to a humidity generator, the ROTRONIC humidity standards offer a reliable and convenient way to generate a reference humidity condition during probe calibration. The ROTRONIC standards are shipped with an internationally accepted certificate, meeting in full the requirements of quality systems such as ISO 9000. The standards are available in boxes of 5 glass ampoules of the same value, and can be stored indefinitely. Standards in the range of 5 to 95 %RH are non-saturated aqueous salt solutions. The 0 %RH humidity standard is made of small granules of a highly porous ceramic. A Material Safety Data Sheet is available for each standard. |
![]() |
ER-15 calibration device: this is a small airtight container that fits on the probe and seals around the humidity sensor. During calibration, a known reference humidity is produced inside the calibration device by means of a humidity standard. |